System for assessing the photobiological hazard of blue light exposure
The measurement system developed by Sedis Light Technology assesses and classifies light sources and luminaires according to the photobiological hazard associated with blue light exposure, in accordance with IEC 62471 and IEC/TR 62778.
To assess blue light hazard, the system uses the alternative method to determine spectral radiance from 300 nm to 700 nm based on spectral irradiance measured with a defined field of view.
Spectral radiance is calculated by dividing the measured spectral irradiance by the solid angle corresponding to the field of view.
Optical configuration
- Spectrometer
- Radiance tube for measuring spectral radiance at the recommended distance of 200 mm, in accordance with IEC/TR 62778:2014
- 2.2 mm diameter aperture (11 mrad field of view)
- 20 mm diameter aperture (100 mrad field of view)
The apertures are interchangeable and their centres are aligned with the centre of the spectrometer’s sensing area.
Measurement procedure
Under IEC/TR 62778:2014, the type of measurement depends on the diameter of the light source.
- Irradiance is measured at 200 mm for sources with a diameter below 2.2 mm
- Radiance is measured at 200 mm for sources with a diameter above 2.2 mm
Most radiance measurements use the 2.2 mm circular aperture fitted to the radiance tube, providing a field of view of 11 mrad.
Test report
The spectrometer connects to a computer and is operated using dedicated software.
Once the light source or luminaire details have been entered and the measurement has been taken, the software processes the data and returns the luminaire’s photobiological risk group within seconds.
The software also generates a PDF test report with the risk group and any applicable labelling instructions.